Pre-installed DO Single-Use-Sensor
For the DO (Dissolved Oxygen) sensor the SUW (Single-Use-Well) is integrated along with assembling of P-SUBs before bagging and subsequent E-beam irradiation. Dry storage capability is 18 month on SUW components stored below 20° and with no UV light exposure.
- CellMembra-500 and CellRetention-500 are both pre-installed with VisiWell (too small vessel for PolarWell)
- CellMembra 3 through 30, CellRetention 3 through 30 pre-installed selectable with either VisiWell or PolarWell
|
VisiWell™ for simple use with the VisiFerm™ optical oxygen sensor range 0-300 % saturated.
The observant viewer do notice that the Stainless-Steel end-cap on VisiFerm is removed and the optical black spot is now visible on the VisiWell plastic part. |
|
|
VisiFerm™ optical DO Re-Usable-Sensor (RUS) is available in the analogue version (short) and the digital ARC version (long). Here shown model 120 and model 225 and both with VP8 connector.
|
|
|
|
PolarWell™ for simple use with most brands of Polarographic, Ampheometric principle DO sensors.
|
|
|
Polarographic DO RUS with typical the 4 wire T82 / D4 classical type of connector. An adapter from T82 / D4 to VP12 is available.
Re-Useable-Sensor (RUS) of relevant use / appear as standard when installed in the PolarWell:
Remember to remove O-ring and possible seal rings on Polarographic, Ampheometric DO RUS. O-rings and/or plastic washer will prohibit correct sensor installment. |
NO INFLUENCE on sensor response time !