Improved cultivation with measured impeller power
Impellers to be found on - https://cercell.com/support/cellvessel-details/impeller-available/
Rotor Watt consumption (W) measured with high precision by Cilix - https://cronus-pcs.com/products/themis-motor-drive
Conditions: 2 liter of tap water at 20 °C in 3,200 ml CellVessel SUB, Head-Plate-Drive (HPD), vessel with 3 baffels, with 3 down tubes (deep tubes or sensors).
Impeller photo |
Impeller name |
300 rpm |
400 rpm |
Kidney
OD 60 mm
3 blades
45 °angle
|
0.2 W
1.9 Np
|
0.4-0.5 W
2.0 Np
|
|
Kidney
OD 60 mm
3 blades
30 °angle
|
0.1 W
1.1 Np
|
0.2-0.3 W
1.1 Np
|
|
Pitch
OD 60 mm
3 blades
45 °angle
|
0.1-0.2 W
1.4 Np
|
0.3-0.4 W
1.4 Np
|
|
Pitch
OD 60 mm
3 blades
30 °angle
30 °angle |
0.1 W
1.0 Np
|
0.2 W
0.9 Np
|
Conditions: water at 20 °C in 3,200 ml CellVessel SUB, Head-Plate-Drive (HPD), vessel without baffles, with 3 down tubes.
Impeller photo |
Impeller name |
300 rpm |
400 rpm |
Kidney
OD 60 mm
3 blades
45 °angle
|
0.2 W
1.8 Np
|
0.4-0.5 W
1.9 Np
|
|
Kidney |
0.1 W
1.0 Np
|
0.2 W
1.0 Np
|
|
Pitch |
0.1-0.2 W
1.3 Np
|
0.3 W
1.3 Np
|
|
Pitch |
0.1 W
0.9 Np
|
0.2 W
0.8 Np
|
Conditions: 2 liter of water at 20 °C in 3,200 ml CellVessel SUB, Head-Plate-Drive (HPD), vessel with/without baffels, with 3 down tubes, impellers mounted on a vertical axle extending down from head plate with no bearing seals for lowest possible mechanical friction.
Impeller photo |
Impeller name |
300 RPM no baffels |
400 RPM no baffels |
300 RPM with baffels |
400 RPM with baffels |
Kidney OD 60 mm 3 blades 30 °angle |
0.1 W 0.9 Np
|
0.1-0.2 W 0.8 Np
|
0.1 W 1.3 Np |
0.2 W |
|
Kidney OD 60 mm 3 blades 45 °angle |
0.1-0.2 W 1.8 Np
|
0.3-0.4 W 1.7 Np |
0.2 W 2.1 Np
|
0.4 W 1.8 Np |
|
Kidney OD 50 mm 3 blades 30 °angle |
0.1 W 1.5 Np |
0.1 W 1.1 Np
|
0.1 W 1.6 Np
|
0.1 W 1.1 Np |
|
Torpedo without impeller 72mm |
0.1 W 0.15 Np
|
0.1 W 0.15 Np
|
0.1 W 0.2 Np
|
0.1 W 0.15 Np |