Select Thalia and SepraPor info

Which SUP volume to insure sufficient velocity

Which ratio between SepraPor and Thalia to select ? The entire P-SUB ATF portfolio follow this chart.

 

OD mm

total

total

HFF

HFF

Thalia stroke

suggested

effective

SepraPor

diameter

length

length

surface

hold-up

volume

Thalia

ratio

part #

mm

mm

mm

cm2

cm3

cm3

 

 

C012-XX1X

9.5

300

 

106

3

25

TM50

8

C024-XX1X

9.5

 

600

220

6.1

25

TM50

4

C112-XX2X

19.1

300

 

431

13

50

TM80

4

C124-XX2X

19.1

 

600

932

26

50

TM80

2

C212-XX2X

25.4

300

 

938

26

50

TM80

2

C224-XX2X

25.4

 

600

1949

54

200

TM100

4

C312-XX3X

31.8

300

 

1224

35

200

TM100

6

C324-XX3X

31.8

 

600

2543

70

200

TM100

3

C412-XX3X

51.8

300

 

4245

122

500

TM140

4

C424-XX3X

51.8

 

600

8817

244

500

TM140

2

General recommendations of flow per lumen: ml/min/lumen/m – 16-25

SUP = Single-Use-Pump = Thalia
HFF = Hollow-Fiber-Filter = SepraPor
ATF = Alternating-Tangential-Flow