0,4 - 2 liter Perfusion-SUB

Medium size SUB for perfusion integrating HFF, SUS and O-SUP

Missing pre-assembled Single-Use-Bioreactor for perfusion? Take advantage of the Pulsating-Tangential-Flow (PTF) concept integrating Single-Use-Bioreactor (SUB), the Clio diaphragm One-way-Single-Use-Pump (O-SUP) with single-use Hollow-Fiber-Filter (HFF) and Single-Use-Sensor's (SUS).

From left: 1. – WORLD FIRST 100 % pre-assembled and single-use assembly, 2. Packed, pre-sterilized and ready to use Single-Use-Bioreactor for perfusion, 3. Rendered assembly incl Single-Use-Pump all ready to use.

CellMembra integrates SUB / O-SUP / HFF / SUS in one package:

  • 3,2 liter Vessel Volume (VV) with 5 x PG 13,5 ports supplied by CerCell
  • 0,4 – 2,0 liter Working Volume (WV)
  • Integrated Single-Use-Sensor - OneFerm or FermProbe pH and VisiWell or PolarWell for standard DO Re-Usable-Sensor's
  • 3 free PG 13,5 port for other sensor's
  • O-SUP is Clio-100 with fully controlled and measured broth volume and velocity - see O-SUP spec 
  • Thermal control either electrical or with water – available for 48 VDC, for 110 VAC, for 230 VAC - or for water
  • A. Spectrum HFF with 0,2 µm pore, 14 x 1,0 mm straw/lumen, 290 cm2 area, Luer-Lok side ports, (T06-P20U-10-N)
  • B. Spectrum HFF with 0,2 µm pore, 75 x 1,0 mm straw/lumen, 980 cm2 area, 3/4" TC side ports, (S04-P20U-10-N)

Clio-100 diaphragm O-SUP controlled by:

  • The ultra compact Clotho Drive Unit available in both a single channel and a dual channel version
  • If you don't have 3-4 Bar pressure and good vacuum supply available in your lab -  then add one or two of the Alagonia pumps and one Sarpedon reservoir

From left: 1. – Working presentation on ESACT in Lausanne. 2. – Incl Clio-100 O-SUP and the support foot which also holds the red laser sensor. 3. – One example of SUB supporting hardware installation showing Clotho, Alagonia, Sarpedon.

SUS - Single-Use-Sensor’s to choose from:

  • One port for either PolarWell or VisiWell Single-Use-Well (SUW) for 220 mm RUS
  • One port for OneFerm or FermProbe pH Single-Use-Sensor (SUS) S8 socket, classical signal
  • PICO bio-mass sensor optional installed horizontal at bottom
  • One port for thermo couple
  • Reduce vastly or eliminate components sterilized in conventional autoclave

Other features for the 3,2 litre P-SUB:

  • Fully assembled Perfusion-SUB packed in dual film bags and precision irradiated - ready to use
  • Limited need for assembling the perfusion concept in LAF bench – less contamination risks - bio-mass SUS is also possible
  • Three baffles of width 10 mm x thickness 4 mm – if you have other requirements – tell us
  • Impeller: Kidney impeller, OD 60 mm, 3 blades, 30 degrees clock wise rotation and up flow mounted 50 mm from the bottom – if you have other requirements – tell us
  • Pumped volume and velocity measured accurately and fully controlled – no guessing


More information about CellMembra P-SUB to be found under Support