Pre-installed DO Single-Use-Sensor
For the DO (Dissolved Oxygen) sensor the SUW (Single-Use-Well) is integrated along with assembling of P-SUBs before bagging and subsequent E-beam irradiation. Dry storage capability is 18 month on SUW components stored below 20° and with no UV light exposure.
- CellMembra-500 and CellRetention-500 are both pre-installed with VisiWell (too small vessel for PolarWell)
- CellMembra 3200 through 13000, CellRetention 3200 through 13000 pre-installed selectable with either VisiWell or PolarWell
VisiWell™ for simple use with the VisiFerm™ optical oxygen sensor range 0-300 % saturated.
The observant viewer do notice that the SS end-cap on VisiFerm is removed and the optical black spot is now visible on the VisiWell. |
|
VisiFerm™ optical DO Re-Usable-Sensor (RUS) is available in the analogue version and the digital ARC version. Here shown model 120 and model 225 and both with VP8 connector. (even the model 225 sensor is named 225 the actual length is 215 mm!)
|
|
|
PolarWell™ for simple use with most brands of Polarographic, Ampheometric principle DO sensors.
|
|
Polarographic DO RUS with typical the 4 wire T82 / D4 classical type of connector. An adapter from T82 / D4 to VP12 is available.
Re-Useable-Sensor (RUS) of relevant use / appear as standard when installed in the PolarWell:
Remember to remove O-ring and possible seal rings on Polarographic, Ampheometric DO RUS. O-rings and/or plastic washer will prohibit correct sensor installment. |
NO INFLUENCE on sensor response time !